Body-protected areas in New Zealand
Medical and dental room electrical rules in New Zealand
In New Zealand, AS/NZS 3003 is the same joint standard, and the Electricity (Safety) Regulations 2010 require an installation intended for electrical medical devices to comply with it. The amended 2018 edition could be used for new work from 13 November 2025; the 2011 edition is accepted until the close of 12 November 2026, so from 13 November 2026 the 2018 edition is the one that counts. Installation work in a patient area is high-risk prescribed electrical work: the electrical worker issues a Certificate of Compliance, a registered electrical inspector inspects the work and issues a Record of Inspection before it is connected, and you receive an Electrical Safety Certificate. Routine inspection and testing at intervals of no more than 12 months applies as in Australia, and practices accredited to the RNZCGP Foundation Standard also have each RCD’s test button pressed every 6 months and recorded.
NZ
High-risk work: inspected and recorded
Regulator: WorkSafe New Zealand (Energy Safety)
Other states
Rules differ by state. Pick yours.
What to do as the practice
- Tell your electrical worker in writing which rooms are patient areas and their boundaries.
- Ask who the registered electrical inspector will be and when the inspection is booked.
- Keep the CoC, the Record of Inspection, the ESC and the AS/NZS 3003 test results with your compliance files.
- Book the 12-month routine test, and have the room tested before any alterations.
- Search the practice’s address on WorkSafe’s high-risk database and check every body-protected room is there.
- If any RCDs sit outside a designated body-protected area, write a transition plan with dates and a budget, and test those RCDs every 6 months until the work is done.
Check a room · Classification record template · Changing an existing room
Testing, everywhere
AS/NZS 3003:2018 (including Amendment 1:2019) requires routine inspection and testing of body-protected areas at intervals not exceeding 12 months. The date of the last test goes on the label on the green sign in the room. The room itself is built to the same rules in New Zealand as in Australia: the rules, one by one.
If the regulator checks
WorkSafe New Zealand’s energy safety team audits electrical work, scheduled, unscheduled or online, chosen by risk, complaints and incidents; the outcome can be an infringement notice or prosecution. Patient-area work is high-risk, so it is also recorded on the public high-risk database within 20 days of the Record of Inspection.
For workplaces, WorkSafe says testing and tagging isn’t mandatory; what the law requires is that equipment is electrically safe and kept that way.
Have these ready
- The Certificate of Compliance, Electrical Safety Certificate and Record of Inspection for every job (keep them at least 7 years)
- The AS/NZS 3003 test results, and a current date on each room’s label
- Your 6-monthly RCD test-button record, if you’re accredited to the RNZCGP Foundation Standard
- Your classification record, and any transition plan
Day-procedure clinics and private hospitals
HealthCERT, at the Ministry of Health, certifies providers of hospital care under the Health and Disability Services (Safety) Act 2001, against Ngā Paerewa NZS 8134:2021. The Act defines hospital care by stays of 24 hours or longer, so check with HealthCERT whether a day-only clinic needs certification.
Questions New Zealand practices ask
From the RNZCGP Foundation Standard, its 2026 scenarios and FAQs, and WorkSafe New Zealand.
- What does RNZCGP accreditation ask for on body-protected areas?
Foundation Standard indicator 11.6 covers residual current devices and body-protected areas. If mains-powered medical equipment is used on… - No body-protected area yet? What goes in a transition plan (New Zealand)
Since RNZCGP rewrote Foundation Standard indicator 11.6 in October 2025, a practice using mains-powered equipment on patients outside a… - Does an electric examination bed need a body-protected area?
Not on its own, in New Zealand’s college guidance. RNZCGP’s electrical safety scenarios (June 2026) say a GP consult room with an electric… - Battery, plug-in and laptop ECGs: which need a body-protected area?
It turns on one thing: whether the patient is connected to something plugged into the wall while the leads are on. New Zealand’s RNZCGP… - How do I check a room on WorkSafe’s high-risk database?
In New Zealand, installation work for electrical medical devices is high-risk prescribed electrical work, and it is recorded on WorkSafe’s… - What if two electricians give different advice?
It happens, usually over whether a room is a patient area or how much of an existing room has to be upgraded. Ask each of them to explain… - Which edition of AS/NZS 3003 applies to my room?
The edition in force when the room was wired. AS/NZS 3003:2018 was published on 26 March 2018 and reissued with Amendment 1 in July 2019;… - Is the body-protected area test every 6 months or every 12 months?
Every 12 months at most. AS/NZS 3003:2018 requires routine inspection and testing at intervals not exceeding 12 months, which is why the… - Can a portable 10 mA RCD replace a body-protected area?
Not as a substitute for wiring the room. A body-protected area is a property of the installation: the 10 mA RCDs in the room, one circuit…
Need an electrician in New Zealand?
Looking for listed electricians in New Zealand…
Ask us to find an electrician in New Zealand
Tell us about the job. We check licences and patient-area experience before recommending anyone, and we’ll send you what to ask whoever you use in the meantime.
Do you do this work in New Zealand?
Electricians and fit-out builders with AS/NZS 3003 experience can be listed here. Licence checked first, free while the directory is new.
Get listedListings are free and licence-checked. How listings work.
Sources for New Zealand
What this page relies on. Every link is re-read monthly; the footer shows when.
Pages we read
- WorkSafe New Zealand: electricity
- WorkSafe New Zealand: changes to the Electricity (Safety) Regulations, new cited standards
- Electricity (Safety) Regulations 2010
- WorkSafe NZ: audits and inspections
- Ministry of Health: certification of health care services
- RNZCGP Foundation Standard 11.6: residual current devices and body-protected areas
- EWRB: AS/NZS 3003:2018 free to licensed workers (standard of the month, August 2026)
- RNZCGP guidance sheet: technical requirements of a body protected area (September 2025, PDF)
- RNZCGP: electrical safety scenarios in general practice (June 2026, PDF)
- RNZCGP: FAQs on body protected areas (July 2026, PDF)
- WorkSafe New Zealand: what is the Electricity and Gas High-risk Database
Sources checked 1 November 2026. Rules change: always confirm with WorkSafe New Zealand (Energy Safety). Spotted something out of date? Tell us. Changes are logged on the Updates page.