What has changed in the body-protected area rules

Every change to AS/NZS 3003:2018 (including Amendment 1:2019), the Wiring Rules and each state’s sign-off rules that matters to a practice, dated, with what it means. Sources re-checked monthly; the last check was 1 November 2026.

Rule-change emails

One email when something changes for your state. No newsletters, nothing else.

Unsubscribe link in every email. Privacy.

2026

This site

Every state page now covers what happens if the regulator checks, and licensing for day-procedure clinics

What changed. Each state and territory page, and the New Zealand page, now says who audits or inspects electrical work after it is done and what they can do about defects, which records to have ready, and the workplace safety switch testing the local regulator expects. A second section names the body that licenses private hospitals and day-procedure centres there and the design rules it applies; Queensland’s building code part for these facilities (MP 5.5) sets minimum numbers of body-protected outlets. Tasmania’s regulator is renamed: CBOS moved into Consumer Affairs Tasmania on 1 October 2026.

Where it applies. Every state page and the New Zealand page.

What it means for a practice. Nothing to do, unless you run a day-procedure clinic or are expecting an audit.

This site

New: a three-question room check

What changed. The room check asks the kind of practice, what is used on the patient in the room, and the state, then shows whether a room like it is usually a body-protected area, how it is signed off in that state, and what to do next. It uses the same practice and state information as the rest of the guide, and the practice still makes the final call.

Where it applies. bpea.com.au/room-check, linked from the home page, every practice page and every state page.

What it means for a practice. Nothing to do. Use it when you are deciding a room.

New Zealand

A New Zealand section: on the map, and six answers for NZ practices

What changed. New Zealand is now on the map beside Australia. Its page has a questions section with six new answers drawn from the RNZCGP Foundation Standard 11.6 guidance and its 2026 scenarios and FAQs, and from WorkSafe New Zealand: what accreditation asks for, what goes in a transition plan when there is no body-protected area yet, electric examination beds, battery, plug-in and laptop ECGs, checking a room on WorkSafe’s high-risk database, and what to do when two electricians disagree.

Where it applies. New Zealand medical and dental practices.

What it means for a practice. If your practice is accredited to the RNZCGP Foundation Standard, read the accreditation and transition plan answers before your next assessment.

Source

This site

Corrections after a sweep of about 150 other sources

What changed. A check of regulators, health departments, testing companies, suppliers and accreditation bodies in Australia and New Zealand found no figure on this site contradicted. It did find loose wording, now fixed: the Clipsal recall now lists all eight part families (ML2031VRC10 was missing), goes by date code rather than purchase date, says what the fault is and gives the national recall number; New Zealand’s cut-over is the close of 12 November 2026, so the 2018 edition is required from 13 November 2026, and the regulation that requires AS/NZS 3003 is named; the 6-monthly push-button check is explained as a real, separate step (RNZCGP in New Zealand, the Electrical Safety Regulation 2026 in Queensland workplaces); the consult-room answer notes that hospitals and licensed facilities wire every patient area as body-protected; the 2.3 m exemption carries its condition; certificate names are right for Western Australia and the ACT; and ACC’s allied health contract requirement is on the physio page.

Where it applies. This site.

What it means for a practice. Nothing to do, unless one of these was your question.

This site

bpea.com.au: the numbers at a glance, a printable quick reference, the standard’s full room list, 15 new questions, and page-level dates

What changed. After a sweep of every other guide on the topic: a table of every figure in the standard with its clause on the rules page; a printable quick reference (page and PDF) with the numbers, the pre-inspection list and the 12-month test; the standard’s list of patient-area rooms; the B, BF and CF equipment symbols; the Statement of Conformity; lettering sizes, individual switching, the 10% colour-coding rule and the alteration label; the 2020 RCD recall; New Zealand accreditation evidence; and 15 new question pages (cost, who can test, landlord or tenant, 6 or 12 months, editions, vets, portable RCDs and more). Every page’s footer now shows when it was published and when its words last changed, and the six regulator pages that block automatic reading were read by hand and dated.

Where it applies. This site.

What it means for a practice. Nothing to do.

This site

bpea.com.au: clause references cross-checked, New Zealand deadline added, electricians’ pre-inspection list

What changed. Every clause number on the rules and alterations pages was cross-checked against the standard’s preface and three published guides to it, and several were made more precise (the 2-metre door rule, RCD position, the test-before-alterations rule, repairs). Added the 12 November 2026 date from which New Zealand requires the 2018 edition, a one-minute summary and a pre-inspection list for electricians on the rules page, and Energy Safe Victoria’s position on like-for-like RCD and outlet replacements.

Where it applies. This site.

What it means for a practice. Nothing changes for a practice. Electricians now have a clause-referenced list to check a room against before certifying or booking the inspector.

This site

bpea.com.au: rules page, alterations page, New Zealand page, Updates page and sources on every state

What changed. Added the rule-by-rule page with section references, two more diagrams, the page on changing an existing practice, a classification record template, a New Zealand page, a sources box on every state page, this Updates page, and the About and Terms pages.

Where it applies. This site.

What it means for a practice. The whole guide now names AS/NZS 3003:2018 including Amendment 1:2019, and every state page shows the Act, regulation and certificate it relies on.

Western Australia

WA: Building and Energy publishes guidance on electrical work in medical centres

What changed. Building and Energy’s ‘Information for electrical licence holders’ now includes a page on electrical work in medical centres (updated 21 September 2026), pointing licence holders to AS/NZS 3003 for patient areas.

Where it applies. Western Australia.

What it means for a practice. WA now has regulator guidance you can show an electrician who hasn’t done patient-area work before.

Source

Queensland

Queensland: Electrical Safety Regulation 2026 commences

What changed. The Electrical Safety Regulation 2026 replaced the 2013 regulation on 1 September 2026. The Electrical Safety Office says it introduces no new policy or obligations; the certificate of testing and compliance process continues as before.

Where it applies. Queensland.

What it means for a practice. Nothing to do. Documents that cite the 2013 regulation now refer to the 2026 one.

Source

New Zealand

New Zealand: AS/NZS 3003:2018 free to licensed workers through the EWRB portal

What changed. The Electrical Workers Registration Board made AS/NZS 3003:2018 its standard of the month and notes that New Zealand licensed electrical workers can read it free through the EWRB online licensing portal.

Where it applies. New Zealand.

What it means for a practice. Your electrical worker or inspector has no cost barrier to reading the current edition.

Source

New South Wales

NSW: every CCEW must now go through eCert

What changed. From 1 July 2026 a Certificate of Compliance for Electrical Work is lodged only through Building Commission NSW’s eCert portal.

Where it applies. New South Wales.

What it means for a practice. Check that the practice’s contact details are right on the job so the eCert copy reaches you, and keep it with the AS/NZS 3003 test results.

Source

Northern Territory

Northern Territory: like-for-like replacements exempt from the certificate of compliance

What changed. NT WorkSafe’s certificate of compliance page, updated 23 April 2026, says a certificate is required for all electrical work except like-for-like replacements, which were exempted in April 2026.

Where it applies. Northern Territory.

What it means for a practice. Replacing a failed outlet or RCD with the same thing in the same place no longer needs a certificate; new or altered patient-area work still does.

Source

2025

New South Wales

NSW: CCEWs can be lodged through the eCert portal

What changed. Building Commission NSW opened its eCert portal for Certificates of Compliance for Electrical Work. Lodging through it was voluntary from 1 December 2025.

Where it applies. New South Wales.

What it means for a practice. Your electrician could start sending your CCEW through eCert; paper and PDF certificates were still accepted until July 2026.

Source

New Zealand

New Zealand regulations cite the amended AS/NZS 3003:2018

What changed. From 13 November 2025 the Electricity (Safety) Regulations 2010 cite AS/NZS 3003:2018 including Amendment 1:2019 for installations used with electrical medical devices, as part of an update to about 440 cited standards. New work can use it from that date; the 2011 edition is accepted until the close of 12 November 2026, so from 13 November 2026 the 2018 edition must be used, and work started under the old edition has to be finished by then.

Where it applies. New Zealand medical and dental premises.

What it means for a practice. Work in a New Zealand patient area is now measured against the same amended edition used in Australia. Anything done after the close of 12 November 2026 must be to the 2018 edition. Existing installations can keep being used and repaired if they are not unsafe.

Source

New Zealand

New Zealand: RNZCGP rewrites its practice standard guidance on RCDs and body-protected areas

What changed. Criterion 11.6 of the Foundation Standard now treats AS/NZS 3003 as legally mandated under the Electricity (Safety) Regulations 2010, asks practices to test their RCDs in-house every 6 months and keep records, and sets out the evidence for accreditation and the transition plan for practices with RCDs outside a designated area. Earlier guidance that treated a body-protected area as recommended is marked outdated.

Where it applies. New Zealand general practices.

What it means for a practice. If you are accredited under the Foundation Standard, keep 6-monthly RCD test records and, if any RCDs sit outside a designated area, a written transition plan.

Source

2020

Victoria

Victoria: Energy Safe Victoria alert on recalled Clipsal 10 mA RCDs

What changed. Schneider Electric recalled a range of Clipsal 10 mA RCD socket-outlets and service panels on 29 October 2020 (part numbers beginning 2025RC10, 2031RC10, 2031VRC10, C2025RC10, ML2025RC10 and ML2031RC10) bought between 1 July 2017 and 15 October 2020. Energy Safe Victoria published the alert, noted that like-for-like replacement of a single RCD or outlet is non-prescribed work, and warned that some RCD testers with 0° and 180° switches cannot perform the AS/NZS 3003 tests.

Where it applies. Victoria, and anywhere those products were installed.

What it means for a practice. If your rooms were fitted out between mid-2017 and late 2020, ask your tester to check the RCD part numbers at the next 12-month test.

Source

2019

AS/NZS 3003

Amendment 1 to AS/NZS 3003:2018 issued

What changed. Corrections and clarifications to the 2018 text. The standard was reissued incorporating the amendment, and the current edition is cited as AS/NZS 3003:2018 including Amendment 1:2019.

Where it applies. Every body-protected room in Australia and New Zealand.

What it means for a practice. Make sure your electrician and inspector are working from the amended edition, not the original 2018 print.

Source

2018

Wiring Rules

AS/NZS 3000:2018 Wiring Rules published

What changed. The general wiring rules that every installation follows. AS/NZS 3003 sits on top of them: the isolator for fixed equipment, RCD protection on ordinary circuits and the earthing a body-protected area uses all come from here. The 2018 edition has been amended since (2020, 2021 and 2023).

Where it applies. All electrical installations, Australia and New Zealand.

What it means for a practice. Nothing specific to body-protected areas, but your electrician works to both standards at once.

Source

AS/NZS 3003

AS/NZS 3003:2018 published, replacing the 2011 edition

What changed. The current edition of the patient areas standard. Compared with 2011: a room is classified from the procedures carried out in it; the practice (the ‘responsible organisation’) must document which areas are patient areas and how each is classified; any patient area that isn’t cardiac-protected must be body-protected; fixed equipment of 2.0 kW or more in a patient area means the whole area has to conform; the distance for protected outlets outside the entrance came down to 2.0 m; new clauses cover USB and other extra-low-voltage charging outlets, and alterations and additions; signs moved to Section 7 and routine testing to Section 8.

Where it applies. Every body-protected room in Australia and New Zealand.

What it means for a practice. Any guide, quote or fit-out spec that still cites AS/NZS 3003:2011 (or 2003) is out of date. Ask for work to AS/NZS 3003:2018.

Source

2011

AS/NZS 3003

AS/NZS 3003:2011 published

What changed. Replaced the 2003 edition. This is the edition many older guides and contractor websites still quote, which is why their clause numbers do not always match the 2018 edition used on this site.

Where it applies. Australia and New Zealand.

What it means for a practice. A room wired to the 2011 edition stays compliant with it until it is altered; any alteration is done to the current edition.

2003

AS/NZS 3003

AS/NZS 3003:2003 published

What changed. Replaced the 1999 edition; itself replaced in 2011.

Where it applies. Australia and New Zealand.

What it means for a practice. Historical. Rooms signed off under it stay compliant with it until they are altered.

1999

AS/NZS 3003

AS/NZS 3003:1999 published

What changed. An earlier joint Australian and New Zealand edition of the patient areas standard, replaced in 2003.

Where it applies. Australia and New Zealand.

What it means for a practice. Historical. A room wired to an older edition and never altered stays on that edition until it is altered.

Monthly checks

On the first of each month the pages’ sources are re-read: every regulator page and piece of legislation cited on the state pages, Building and Energy’s medical centres guidance, WorkSafe New Zealand, and the Standards Australia and Standards New Zealand listings for AS/NZS 3003 and AS/NZS 3000. A ‘no change’ month is logged here as well, so you can see the checking happened.

  • Sources checked: 2 changes under review 47 sources re-read. 2 changes that may matter: NSW (AI check failed ('NoneType' object has no attribute 'group'); flagging for a per); WA (AI check failed ('NoneType' object has no attribute 'group'); flagging for a per). Being checked by a person before the pages change; a dated entry appears above when it is published.
  • Sources checked, nothing changed 25 sources re-read: every regulator page, Act and regulation cited on the state and New Zealand pages, and the Standards Australia and Standards New Zealand listings for AS/NZS 3003 and AS/NZS 3000. No changes that affect the guide. Could not read 6: NSW: could not read Gas and Electricity; QLD: could not read WorkSafe Queensland: issuing certificates of compliance; QLD: could not read WorkSafe Queensland: Electrical Safety Regulation 2026; TAS: could not read CBOS: notification of electrical work; ACT: could not read ACT Government: electrical inspections; NT: could not read NT WorkSafe: certificate of compliance.

How a change gets onto the site

  • A source changes. The difference is read and summarised.
  • If it affects who certifies, inspects or tests, or what a room needs, a rewrite of the affected page is drafted.
  • A Victorian registered electrical contractor reads the source and the draft, then publishes it. Nothing changes on the site unread.
  • The change is logged above and emailed to practices in that state who asked to hear.