Patient area classification record

The written record of which rooms are patient areas, how each is classified, which rooms are not and why, and every change. AS/NZS 3003:2018 (including Amendment 1:2019) asks the practice to keep it on the premises; Victoria’s regulations and Western Australia’s guidance expect it to be given to the electrician in writing before work starts.

Practice
Address
Responsible person
Record last updated
Room (name or number)What happens there (procedures)Mains-powered equipment applied to patientsClassificationIf not a patient area, whyBoundaries (walls and doors)Decided by and dateSign and test label fittedLast test dateNext test due

Filling it in

  • Classification is one of: body-protected, cardiac-protected, or not a patient area. If mains-powered medical equipment is applied to a patient in the room, it is a patient area.
  • Not a patient area needs a reason, for example ‘consultation only, no mains equipment used on patients’. Write it down; that is what the record is for.
  • Boundaries are full-height walls and doors. Note an ensuite as part of the area.
  • Decided by is the practice, naming who was consulted: clinicians, equipment suppliers, the electrician.

Keeping it current

  • Update it whenever a room’s use changes, equipment is added or an area is altered, and date the change.
  • Give a copy to the electrician in writing before any work, and to the inspector or certifier when they test.
  • Keep it with the certificates and the AS/NZS 3003 test results: auditors, insurers and the next electrician will ask for all three together.
  • Set the 12-month reminder from the earliest ‘next test due’ date on the sheet.

From bpea.com.au · a template, not a certificate. The record is the practice’s own document under Clause 2.2.1 of AS/NZS 3003:2018 (including Amendment 1:2019). Information, not advice: your electrician and your state regulator decide what applies to your rooms.